Glucose measurement test, Comfort to live with confidence
Glucose measurement systems Max features an elegant and intuitive interface that provides measurement security. This sleek, dark-designed system features a wide range of HCTs, automatic coding, status markers, and ketone alerts to keep any user informed and in control.
Full compliance with EN ISO 15197: 2015 and ISO 15197: 2013
-
Wide range of HCT
Suitable for neonatal and anemic patients. -
Ketone alert
A warning is displayed when ketone levels need to be checked. -
Status markers
Be mindful of exercise times and meals.
-
Complete blood sugar management *
Add notes, photos, and status indicators for more comprehensive entries. -
Personalization*
Set blood glucose goals and customize your measurement program. -
Intuitive statistics panel *
Useful data is clearly presented.
* Bluetooth connectivity available only with Max Plus. Transfer results wirelessly to the Rightest CARE app.
Glucose variability:
Hyper

Normal

Hypo


Markers:

- Midnight
- Before meal
- After meal
- Wake up
- Exercise
- Bedtime
How does it work

1.Insert the strip into the meter socket. Calibration will happen automatically

2.Tap a blood sample against the sample inlet port of the strip.

3.After 5 seconds, the test result will appear.

4.Rotate and pull up to remove the test strip
For more information on using the device, see the instructions for use.
- Rightest Max Catalog / Rightest Max Plus Catalog
- Getting Started Guide
- Lancing Device Instructions.
- Glucose Test Strip Instructions.
Specifications Max / Max Plus
Test strip
Coding
Enzyme / Measurement technology
Blood sample source
Minimum sample volume
HCT range
Measuring range
Meter dimensions (L x W x H)
Display area (L x W)
Weight (with battery)
Power supply
Memory capacity
Operating temperature
Operating relative humidity
Meter storage conditions
Test strip storage conditions
Bluetooth version (For GM700S)
Tira Max
Automatic coding
Sensor GDH / electroquímico
Capillary, venous, arterial and neonatal whole blood
0,75 µL
10 – 70%
0,6 – 33,3 mmol / L (10 – 600 mg / dL)
82,0 x 50,0 x 15,5 mm (3,23 x 1,97 x 0,61 pulgadas)
40,7 x 40,2 mm (1,60 x 1,58 pulgadas)
59,0 ± 5 g
1 × batería CR2032 para Max | 2 × baterías CR2032 para Max Plus
500 resultados de pruebas con fecha y hora
6 – 44 ° C (43 – 111 ° F)
10 – 90%
-10 – 60 ° C (14 – 140 ° F)
A 4 – 30 ° C (39 – 86 ° F) y por debajo del 90% de humedad relativa
4.0
Accuracy
Precision
For glucose concentrations < 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),
95% of measured glucose values within ± 15% of reference measurements.
For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%