Never Lose Track Again

The GM700S is a premium meter featuring FAD-GDH reagent technology to minimize oxygen interference and keep you updated with results you can trust. Accepts a wide HCT range and comes with helpful meal markers to keep you as informed as possible.

Full compliance with EN ISO 15197: 2015 and ISO 15197: 2013

*Bluetooth connectivity available with GM700SB only. Transfer results wirelessly to the Rightest CARE App.

Markers:

Rightest Bluetooth BGMs

Unlock the full potential of seamless data transfer
  • Simple Storage
  • Immediate Upload
  • Glucose Analysis

Synchronize with the Rightest CARE app

Simplified glucose management

How does it work

1.Insert the strip into the meter socket. Calibration will happen automatically

2.Tap a blood sample against the sample inlet port of the strip.

3.After 5 seconds, the test result will appear.

4.Rotate and pull up to remove the test strip

Especificaciones GM700S / GM700SB

Test strip

Coding

Enzyme / Measurement technology   

Blood sample source 

Minimum sample volume

HCT range    

Measuring range

Meter dimensions (L x W x H)        

Display area (L x W)         

Weight (with battery)      

Power supply

Memory capacity          

Operating temperature

Operating relative humidity

Meter storage conditions          

Test strip storage conditions        

Versión de Bluetooth (para GM700SB )

GS700

Automatic coding

Sensor GDH / electroquímico

Capillary, venous, arterial and neonatal whole blood

0,75 µL

10 – 70%

0,6 – 33,3 mmol / L (10 – 600 mg / dL)

82,0 x 45,0 x 15,5 mm (3,22 x 1,77 x 0,61 pulgadas)

34,0 x 27,5 mm (1,34 x 1,09 pulgadas)

43,0 ± 5 g

1 × batería CR2032

500 resultados de pruebas con fecha y hora

6 – 44 ° C (43 – 111 ° F)

10 – 90%

-10 – 60 ° C (14 – 140 ° F)

A 4 – 30 ° C (39 – 86 ° F) y por debajo del 90% de humedad relativa

4.0

Accuracy

 

 

                                                                                                               

 

Precision

 

For glucose concentrations < 5.5 mmol/L (100 mg/dL),

95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements.

For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL),

95% of measured glucose values within ± 15% of reference measurements.

For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL)

For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5%